Waferpedia

Zeiss

Supra 55

Metrology & InspectionZeiss Supra 55 family
Research Quality: 70% complete

The Zeiss Supra 55 is a field-emission scanning electron microscope (FE-SEM) that uses a Schottky thermal field emission source to achieve high-resolution imaging. The Supra 55 provides an ultimate imaging resolution in the range of 2 nm and an accelerating voltage range of 0.2 to 30 kV.[1][2]

Supra 55 — Inventory photo
Fig. 01Supra 55Inventory photo[3]
  • Plate 01Zeiss Supra 55 Training Video Part 1 Review

    David FreyWatch on YouTube

Power

0.2 to 30 kV[2]

Vacuum

Adixen rough pump (for airlock)[3]

Optics

Detectors:[3]

What it is

The Zeiss Supra 55 is a scanning electron microscope used for metrology and inspection. The Zeiss Supra 55 belongs to the field-emission SEM class and is used for high-resolution surface imaging and characterization.[1][2][4][5]

How it works

The Zeiss Supra 55 uses multiple detectors for secondary-electron and backscatter signals. The Zeiss Supra 55 can also support cathodoluminescence detection and electron-beam-lithography-related use when configured for those functions.[2][1]

The Zeiss Supra 55 VP operates in high vacuum and variable-pressure modes. The Zeiss Supra 55 VP includes a load-lock system for specimen exchange without venting the main chamber.[4][5]

Where it fits in the process flow

The Zeiss Supra 55 is used alongside analytical methods when elemental or contrast information is needed. The Zeiss Supra 55 can support electron-beam-lithography-related work in facilities that configure it for that use.[1][2][4][5]

Applications

The Zeiss Supra 55 is used for high-resolution imaging of surface topography. The Zeiss Supra 55 is used for secondary-electron imaging, backscatter imaging, and composition-sensitive contrast when fitted with the relevant detectors.[2][1]

The Zeiss Supra 55 is used for materials characterization. The Zeiss Supra 55 is also used for nanomanipulation and force measurement in setups that include those accessories.[2]

The Zeiss Supra 55 VP is used for samples with a wide range of conductivity. The Zeiss Supra 55 VP is used for solid-sample compositional analysis when equipped with energy-dispersive spectroscopy.[5][4]

What do the numbers mean?

Power & electrical3

Accelerating Voltage
0.2 to 30 kV[2]
Accurate?
Accelerating Voltage (VP variant)
0.1 to 30 kV[4]
Accurate?
Acceleration voltage
0.1 - 30 kV[4]
Accurate?

Vacuum & pumping10

Adixen rough pump (for airlock)[3]
Accurate?
Edwards RV12 Oil Rough pump (Chamber)[3]
Accurate?
Imaging Resolution (High Vacuum, 15 kV)
1.0 nm[4]
Accurate?
Imaging Resolution (High Vacuum, 1 kV)
1.7 nm[4]
Accurate?
Imaging Resolution (High Vacuum, 0.2 kV)
3.5 nm[4]
Accurate?
Variable Pressure Range
2 to 133 Pa, adjustable in steps of 1 Pa[4]
Accurate?
Pumps
Adixen rough pump (airlock), Edwards RV12 oil rough pump (chamber)[3]
Accurate?
Variable pressure range
2 - 133 Pa, adjustable in steps of 1 Pa[4]
Accurate?
Accessory
Adixen rough pump (for airlock)[3]
Accurate?
Accessory
Edwards RV12 Oil Rough pump (Chamber)[3]
Accurate?

Optics & imaging9

Detectors:[3]
Accurate?
Detectors[3]
Accurate?
Imaging Resolution (VP mode, 30 kV)
2.0 nm[4]
Accurate?
Resolution (general range)
2 nm[1]
Accurate?
Magnification Range
12 - 900,000x[4]
Accurate?
Detectors
InLens SE, E-T (SE2), AsB (backscattered), CL (cathodoluminescence), STEM, EDS, EBSD[3]
Accurate?
Electron image resolution
1.0 nm (@ 15 kV), 1.7 nm (@ 1 kV), 3.5 nm (@ 0.2 kV), 2.0 nm (@ 30 kV in VP mode)[4]
Accurate?
Magnification
12 - 900,000x[4]
Accurate?
Detectors
InLens, SE, 4QBS, STEM, CCD PC[3]
Accurate?

Control & software2

PC[3]
Accurate?
Operating System
Windows 7 32-bit with Smartsem v5.05[3]
Accurate?

Configuration & options23

InLens[3]
Accurate?
4QBS[3]
Accurate?
STEM[3]
Accurate?
CCD[3]
Accurate?
Win7 32bit[3]
Accurate?
Smartsem v5.05[3]
Accurate?
Accessory[3]
Accurate?
FJELD Airlock[3]
Accurate?
ATC Chiller[3]
Accurate?
EDAX ELECT SUPER 70mm2E[3]
Accurate?
Electron Source Type
Schottky thermal field emission source (FESEM)[1]
Accurate?
Probe Current
4 pA to 10 nA (20 nA optional)[4]
Accurate?
Sample Stage
5-Axes (X, Y, Z, R, T) Motorized Cartesian[4]
Accurate?
Load-Lock System
FJELD Airlock[3]
Accurate?
Chiller
ATC Chiller[3]
Accurate?
Model
Supra 55[3]
Accurate?
Equipment type
Scanning Electron Microscope (SEM)[3]
Accurate?
Source type
FESEM[1]
Accurate?
Electron source
Schottky thermal field emission source[1]
Accurate?
Probe current
4 pA - 10 nA (20 nA optional)[4]
Accurate?
Accessory
FJELD Airlock[3]
Accurate?
Accessory
ATC Chiller[3]
Accurate?
Accessory
EDAX ELECT SUPER 70mm2E[3]
Accurate?

Vintage & configurations

Documented models & variants

DesignationGenerationVintageChangesSource
Supra 55 VP——Variable-pressure version; sources state high-vacuum and VP modes, with a load-lock system on one source and VP capability up to 1.33 mbar on another.bu.edu[5]
Supra 55 FESEM——Described as a FESEM with a Schottky thermal field emission source.cmdis.rpi.edu[1]
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • ConfigurationAdixen rough pump (for airlock)[3]
  • ConfigurationEdwards RV12 Oil Rough pump (Chamber)[3]
  • Accelerating Voltage0.2 to 30 kV[2]
  • Accelerating Voltage (VP variant)0.1 to 30 kV[4]
  • Imaging Resolution (High Vacuum, 15 kV)1.0 nm[4]
  • Imaging Resolution (High Vacuum, 1 kV)1.7 nm[4]
  • Imaging Resolution (High Vacuum, 0.2 kV)3.5 nm[4]
  • Variable Pressure Range2 to 133 Pa, adjustable in steps of 1 Pa[4]
  • PumpsAdixen rough pump (airlock), Edwards RV12 oil rough pump (chamber)[3]
  • Variable pressure range2 - 133 Pa, adjustable in steps of 1 Pa[4]
  • Acceleration voltage0.1 - 30 kV[4]
  • AccessoryAdixen rough pump (for airlock)[3]
  • AccessoryEdwards RV12 Oil Rough pump (Chamber)[3]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What type of microscope is the Zeiss Supra 55?

The Zeiss Supra 55 is a field-emission scanning electron microscope.[4][5]

Does the Zeiss Supra 55 support variable pressure operation?

The Zeiss Supra 55 VP operates in both high vacuum pressure and variable pressure modes.[4]

Not publicly documented

The following facts about the Supra 55 are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the Supra 55 are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • The control-system platform and OS era of the Supra 55 are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the Supra 55 are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the Supra 55 is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the Supra 55 are on record.

    Answerable by: an OEM parts catalog or a service engineer

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (6)Every fact above is drawn from these public sources
  1. [1]cmdis.rpi.edu — cmdis.rpi.educmdis.rpi.edu
  2. [2]weizmann.ac.il — weizmann.ac.ilweizmann.ac.il
  3. [3]Inventory photo
  4. [4]binghamton.edu — binghamton.edubinghamton.edu
  5. [5]bu.edu — bu.edubu.edu
  6. [6]BEEM: Zeiss Supra 55 VP — tuhh.detuhh.de
Was this page accurate?

Last updated Oct 10, 2026.

Compare with similar tools

View all →