| Attribute | 1050 UV-VISPerkin Elmer | LTQ Orbitrap XL MassThermo Fisher |
|---|---|---|
| OEM | Perkin Elmer | Thermo Fisher |
| Category | Precision Optics | Precision Optics |
| Wafer size | — | — |
| Process node | — | — |
| Introduced | — | — |
| Production run | — | — |
| Lifecycle | — | — |
| Lifecycle milestones | — | — |
| Control system | — | — |
| Generation | — | — |
| Family | Perkin Elmer 1050 UV-VIS | Thermo Fisher LTQ Orbitrap XL Mass |
| Specifications | ||
| Asset ID | 82948[1] | — |
| Last Confirm Date | August 23, 2024[1] | — |
| Manufacturer | Perkin Elmer[1] | — |
| Model | 1050[1] | LTQ Orbitrap XL[2] |
| Instrument type | Spectrometer[1] | Hybrid mass spectrometer[4] |
| Use | Analyzing ultraviolet and visible light absorption spectra[1] | — |
| Max Resolution | — | 100,000[2] |
| Configuration | — | IonMax ESI Source[2] |
| Make | — | Thermo Scientific[2] |
| Ion source | — | IonMax ESI Source[2] |
| Mass spectrometer dimensions | — | 141.4 × 87 × 146.3 cm (H × W × D)[3] |
| Mass spectrometer weight | — | ~600 kg[3] |
| Power requirements | — | 230 Vac ±10% 3 phase, 16 Amps, 50/60 Hz, with earth ground for the instrument[3] |
| Gas supply | — | One high purity (99.5% pure, flow rate 15 L/min) nitrogen gas supply for the API source; one ultra-high-purity helium gas supply (99.998%) with less than 1 ppm each of water, oxygen, and total hydrocarbons for the linear ion trap[3] |
| Maximum resolution | — | 100,000[2] |
| Resolution (apodized) | — | 60,000 (FWHM) at m/z 400 with a scan repetition rate of 1 second[4] |
| Minimum resolution | — | 7,500 at m/z 400[4] |
| Mass range | — | m/z 50–2000; m/z 200–4000[4] |
| Mass accuracy | — | <3 ppm RMS for 2 h period with external calibration using defined conditions; <2 ppm RMS with internal calibration[4] |
| Dynamic range | — | >10000 between mass spectra; >4000 between highest and lowest detectable mass in one spectrum[4] |
| Collision cell | — | HCD collision cell[4] |
| Recirculating chiller | — | Thermo Scientific ThermoFlex 900[4] |
| Main components | — | Linear ion trap, intermediate storage device (curved linear trap), Orbitrap analyzer, collision cell[4] |
| Resolution range | — | 7,500 to 100,000 at m/z 400[4] |
| Scan functions | — | MS/MS and MSn[4] |
| HCD collision cell | — | Included[3] |
| Power | — | 230 Vac ±10% 3 phase, 16 Amps, 50/60 Hz, with earth ground for the instrument; 120 or 230 Vac single phase with earth ground for the data system; 120 or 230 Vac single phase, 15 Amps, with earth ground for the water chiller[3] |
| Weight | — | ~600 kg[3] |
| Dimensions | — | 141.4 × 87 × 146.3 cm (H × W × D)[3] |
Plan your fab